Apparatus Components & Accessories

rotameter_without_tubes_thumb

Rotameter Unit (Without Tubes)

A housing for Rotameter tubes. The flow control is obtained by needle valves which can be easily distinguished from one another by touch & colour. Easy to read large size scales with fluorescent illumination, allow inside bobbins to be accurately read.

3700 - Accomodates one 230mm tube
3600 - Accomodates two 230mm tubes
3500 - Accomodates three 230mm tube


Rotameter Tube & Bobbin

3601 - Oxygen              0.1 to 10.0 ltr/min
                                   100 to 10000 cc/min   
3602 - Nitrous Oxide     0.2 to 12.0 ltr/min
                                   200 to 12000 cc/min
3612 - Air                    0.1 to 12.0 ltr/min
                                   100 to 12000 cc/min
3613 - Oxygen             0 to 1.0 ltr/min
                                  0 to 1000 cc/min
Dual tapered calibrated tube calibrated at 20OC & 1013mbar abs.
Bobbin Float : Aluminium alloy anodised red, with silver dot for easy observation of rotation. Operation of the tubes can be easily confirmed by checking the location of flowmeter bobbin & by seeing the rotation of the bobbin.

yoke_economy_model_thumb

Yoke - Economy Model

3652 - Oxygen    3653 - Nitrous Oxide
Colour coded plastic cover, other components from stainless steel to avoid rusting, pin indexed for use on anaesthesia trolley.

3650_yoke_compact_model_thumb

Yoke - Compact Model

3650 - Oxygen    3651 - Nitrous Oxide
One piece, brass chromed, robust & pin indexed for use on anaesthesia trolley.

3654_yoke_midget_thumb

Yoke - Midget

3654 - Oxygen for use with F. A. Valve
3655 - Nitrous Oxide
3656 - Oxygen for use with Rao’s Regulator
For use other than anaesthesia trolley. Pin indexed, brass chromed, specially developed for flush type portable cylinders for use with portable anaesthesia apparatus. Ideal for domiciliary use, ambulances, mines, airports, railways and medical disaster situations.

3604_change_over_unit_thumb

Change Over Unit

3604
Normally positioned at the top of the apparatus table. It has a built in emergency oxygen push button and change over facility from open to closed circuit and vice versa.

3606_emergency_oxy_unit_thumb

Emergency Oxygen Unit

3606
Incorporates a non-return valves assembly along with a button for flushing the circuit with oxygen in emergencies.

3603_non_return_valve_thumb

Non Return Valve Assembly

3603
Blows off when pressure exeeds 200cms of water column thus minimising the risk of back flow of gases.

3607_o2_failure_alarm_thumb

O2 Failure Alarm - for Gas Anaesthesia Appartus

3607
A battery operated pnuematically controlled device which activates an audio alarm for about fifteen seconds when oxygen pressure in the system falls to approximately 4.2kg/cm2

4002_goldman_halothane_vapo_thumb

Goldman Halothane Vaporizer - w/o stand

4002
Drawover type, designed to limit the maximum concentration of Halothane vapour to 3% by volume irrespective of the total gas flow through the unit. Adjustment is made by means of a ratchet type control valve. The glass jar is screwed on to the body and can be quickly detached for easy filling. The vaporizer is also suitable for use with Trichloroethylene in low concentrations.

3625_raos_preset_regulator_thumb

Rao's Preset Regulator

3625 - Oxygen    3626 - Nitrous Oxide
For use with anaesthesia apparatus only. Outlets have non-return valves to avoid back flow of gases. Safety valve is incorporated in the unit. Hence no necessity of safety valve in the tubing between the regulator and the rotameter unit.

3629_raos_cut_off_regu_thumb

Rao's Cut Off Regulator

3629
To activate Nitrous Oxide Regulators only when Oxygen Regulators are pressurised, thereby minimising the risk of administering Nitrous Oxide in the absence of Oxygen. Safety valve incorporated in the unit. Hence no necessity of safety valve in the tubing between the regulator & the rotameter unit.

raos_bulk_cylinder_cnversion_thumb

Rao's Bulk Cylinder Conversion Units

3675 - Oxygen    3676 - Nitrous Oxide
For connecting bulk cylinder to the anaesthesia apparatus. The unit comprises of a Regulator, 2 mtrs. High Pressure Tubing with ferrules and a Pin Index Adaptor to fit on the yoke of the apparatus in place of the service cylinder.

bag_mounts_thumbs

Bag Mounts

9022 - 22F x 22M & 22M at Bag end - Plastic
9023 - 22F x 22M & 22F at Bag end - Plastic
9025 - 23M x 23F & 22 F at Bag end - Plastic
The above will not bend or dent when accidentally dropped.
9027 - 23M x 23F & 23 M at Bag end with ON/OFF Lever - Brass chromed.

expiratory_valves_heidbrink_thumb

Expiratory Valves - Heidbrink

9200 - 22M/15F at patient end & 22F at distal end
9201 - 23M/15F at patient end & 23F at distal end
15mm E/T straight connectors can be fitted at the patient end, thus avoid using catheter mount & reducing dead space. Will not bend or dent when accidentally dropped.
9202 - 23M/23F Brass Chromed
A variable tension spring is fitted in the above three expiratory valves to control the degree of rebreathing.

9028_22angle_wye_expi_thumb

22 Angle "Wye" w/o Expiratory Valve

9028
The patient end accepts anaesthesia mask & 15mm tracheal connector.

9050_dr_nawathes_swivel_expiratory_valve_thumb

Dr. Nawathe's Swivel with Expiratory Valve

9050
Patient end accepts anaesthesia mask & 15mm tracheal connector.  Low weight - 50gms - leads to minimization of the pressure strain for patient. Sturdy to withstand rough usage, will not bend or dent when accidentally dropped, needs no servicing with absence of circlips & rubber washers.